BrightLine® CFP/YFP FRET Filter Set

   
CFP/YFP FRET Filter Set
 Optimized for these fluorophores
Cyan: CFP, CyPet, AmCyan
Yellow: YFP, YPet, Venus
Actual measured data from typical filters
Exciter Dichroic Emitter 1
ASCII Data ASCII Data ASCII Data
Emitter 2
ASCII Data
 Set description
This four-filter set is specially designed to achieve high-performance fast FRET (Fluorescence Resonance Energy Transfer) imaging for cyan donor fluorescence and yellow acceptor fluorescence. FRET measurements using wide-field microscopy are challenging – using the best possible filters can make a significant difference. These filter combinations provide excellent brightness and signal-to-noise ratio while minimizing the undesirable donor-acceptor crosstalk. In systems with dual (exciter-emitter) filter wheels, try our CFP/YFP-2X2M-A  dual-band Sedat set for FRET.
Filter Average Transmission Center Wavelength Bandwidth *
 Exciter > 93% 438 nm 24 nm
 Emitter 1 > 93% 483 nm 32 nm
 Emitter 2 > 93% 542 nm 27 nm
  Average Reflection Reflection Band Transmission Band
 Dichroic > 98% 426 - 450 nm 467 - 600 nm
* Bandwidth is the minimum width over which the average transmission exceeds the specified passband transmission; the nominal full-width-at-half-maximum (FWHM) is approximately the Bandwidth + 1% of the Center Wavelength.