This
four-filter set is specially designed to achieve high-performance
fast FRET (Fluorescence
Resonance Energy Transfer) imaging for blue donor fluorescence
and green acceptor fluorescence. FRET measurements using wide-field
microscopy are challenging – using the best possible
filters can make a significant difference. These filter combinations
provide excellent brightness and signal-to-noise ratio while
minimizing the undesirable donor-acceptor crosstalk.
Filter
Average
Transmission
Center
Wavelength
Bandwidth
*
Exciter
>
85%
387
nm
11
nm
Emitter
1
>
93%
447
nm
60
nm
Emitter
2
>
93%
520
nm
35
nm
Average
Reflection
Reflection
Band
Transmission
Band
Dichroic
>
98%
344
- 404 nm
415
- 570 nm
* Bandwidth is the
minimum width over which the average transmission exceeds the specified
passband transmission; the nominal full-width-at-half-maximum (FWHM)
is approximately the Bandwidth + 1% of the Center Wavelength.