Search Semrock's website:

BrightLine® QDLP-A Filter Set

   
QDLP-A Filter Set
 Optimized for this fluorophore
Qdot® 525, 565, 585, 605, 625, 655, 705, 800 Nanocrystals (from Invitrogen's Molecular Probes). Also compatible with Evident Technologies EviTag quantum dots and EviFluor conjugates, antibodies, and proteins.
Actual measured data from typical filters
Exciter Dichroic Emitter
ASCII Data ASCII Data ASCII Data
 Set description
A special long-pass emission filter in this filter set allows simultaneous viewing by eye or with a color camera of samples that are multi-labeled with quantum dot nanocrystals. It takes advantage of the unique fact that all quantum dots share a common UV/blue excitation spectrum while simultaneously exhibiting narrow and specific visible emission spectra. The highly transmitting, deep-blue exciter efficiently excites all quantum dots with no UV light to damage samples or generate undesirable background, while the high and uniform transmission of the long-pass emission filter provides exceptional and balanced brightness. And with the "no burn-out" performance shared by all BrightLine filters, the permanent performance of this set will outlast even your quantum dots!
Filter Average Transmission Center Wavelength Bandwidth *
 Exciter > 90% 435 nm 40 nm
 Emitter > 90% 515 nm (edge) 519 - 700 nm
  Average Reflection Reflection Band Transmission Band
 Dichroic > 98% 327 - 488 nm 515 - 850 nm
* Bandwidth is the minimum width over which the average transmission exceeds the specified passband transmission; the nominal full-width-at-half-maximum (FWHM) is approximately the Bandwidth + 1% of the Center Wavelength.
Copyright © 2008 Semrock, Inc.