QDLP-A
Filter Set |
|
|
 |
| Optimized
for this fluorophore |
| Qdot®
525, 565, 585, 605, 625, 655, 705, 800 Nanocrystals
(from Invitrogen's Molecular Probes). Also compatible with
Evident Technologies EviTag quantum dots and EviFluor conjugates,
antibodies, and proteins. |
|
 |
| Actual
measured data from typical filters |
|
 |
| A
special long-pass emission filter in this filter set allows
simultaneous viewing by eye or with a color camera of samples
that are multi-labeled with quantum
dot nanocrystals. It
takes advantage of the unique fact that all quantum dots
share a common UV/blue excitation spectrum while simultaneously
exhibiting narrow and specific visible emission spectra.
The highly transmitting, deep-blue exciter efficiently
excites all quantum dots with no UV light to damage samples
or generate undesirable background, while the high and
uniform transmission of the long-pass emission filter provides
exceptional and balanced brightness. And with the "no
burn-out" performance shared by all BrightLine filters,
the permanent performance of this set will outlast even
your quantum dots! |
|
 |
| Filter |
Average
Transmission |
Center
Wavelength |
Bandwidth
* |
| Exciter |
> 90% |
435
nm |
40
nm |
| Emitter |
> 90% |
515 nm
(edge) |
519
- 700 nm |
| |
Average
Reflection |
Reflection
Band |
Transmission
Band |
| Dichroic |
> 98% |
327
- 488 nm |
515
- 850 nm |
 |
| * Bandwidth
is the minimum width over which the average transmission
exceeds the specified passband transmission; the nominal
full-width-at-half-maximum (FWHM) is approximately the
Bandwidth + 1% of the Center Wavelength. |
|
 |
|
|