BrightLine® FITC/TXRED-A Full Dual-band Filter Set

   
 
FITC/TXRED-A Full Dual-band Filter Set
 Optimized for these fluorophores
Green:  FITC, GFP, rsGFP, Bodipy, Alexa Fluor 488
Red:
 Texas Red®, mCherry, Alexa Fluor 568 & 594 
Complete list of fluorophores for this filter set
Actual measured data from typical filters
Exciter Dichroic Emitter
ASCII Data ASCII Data ASCII Data
 Set description
This new "full-multiband" filter set allows for simultaneous direct viewing or imaging with a color camera. Popular 2-fluorophore combinations include FITC/Texas Red and GFP/mCherry. For a "Pinkel" version of this filter set (replaces the dual-band emitter with two single-band emitters) see the FITC/TXRED-2X-A set, and for a "Sedat" version of this set (replaces the dual-band emitter and exciter with two single-band emitters and two single-band exciters) see the FITC/TXRED-2X2M-A set
Filter Average Transmission Center Wavelengths Bandwidths *
 Exciter > 90%, 90% 479, 585 nm 38, 27 nm
 Emitter

> 90%, 90%

524, 628 nm 29, 33 nm
  Average Reflection Reflection Bands Transmission Bands
 Dichroic > 95%
> 95%
458 - 499 nm
570 - 600 nm
509 - 541 nm
612 - 647 nm
* Bandwidth is the minimum width over which the average transmission exceeds the specified passband transmission; the nominal full-width-at-half-maximum (FWHM) is approximately the Bandwidth + 1% of the Center Wavelength.