This
new "full-multiband" filter set allows for simultaneous
direct viewing or imaging with a color camera. For a "Pinkel"
version of this filter set (replaces the dual-band emitter
with two single-band emitters) see the Cy3/Cy5-2X-A
set, and for a "Sedat"
version of this set (replaces the dual-band emitter and
exciter with two single-band emitters and two single-band
exciters) see the Cy3/Cy5-2X2M-A
set
Using
a split-camera detection system and need to split your Cy3/Cy5
emission? Try our FF655-Di01-25x36.
Filter
Average
Transmission
Center
Wavelengths
Bandwidths
*
Exciter
>
90%, 90%
534,
635 nm
36,
31 nm
Emitter
>
90%, 90%
577,
690 nm
24,
50 nm
Average
Reflection
Reflection
Bands
Transmission
Bands
Dichroic
>
95%
> 95%
514
- 553 nm
617 - 652 nm
564
- 591 nm
665 - 718 nm
* Bandwidth
is the minimum width over which the average transmission
exceeds the specified passband transmission; the nominal
full-width-at-half-maximum (FWHM) is approximately the
Bandwidth + 1% of the Center Wavelength.