BrightLine Basic™ FITC Long-pass Filter Set

   
FITC-LP01-Clinical Long-pass Filter Set
 Optimized for these fluorophores
FITC and Acridine Orange
Actual measured data from typical filters
Exciter Dichroic Emitter
ASCII Data ASCII Data ASCII Data
 Set description

This long-pass filter set is optimized for all popular clinical tests based on blue excitation and green fluorescence. These include both direct (DFA) and indirect (IFA) immunofluorescent tests utilizing the FITC fluorochrome. Samples will exhibit amazing brightness and superior contrast for easy and accurate identification in test after test. Because it is a long-pass set, for visual applications it offers higher brightness and the ability to observe longer-wavelength (e.g. yellow, orange, and red) fluorescence from counter-stains such as Rhodamine and Evans Blue. For our best-value bandpass filter set, try FITC-A-Basic.

Filter Average Transmission Center (Edge) Wavelength Bandwidth *
 Exciter > 90% 475 nm 28 nm
 Emitter > 90% 515 nm (edge) 519 - 700 nm
  Average Reflection Reflection Band Transmission Band
 Dichroic > 90% 462 - 490 nm 519 - 700 nm
* Bandwidth is the minimum width over which the average transmission exceeds the specified passband transmission; the nominal full-width-at-half-maximum (FWHM) is approximately the Bandwidth + 1% of the Center Wavelength.
 Compare BrightLine Basic filters to soft-coated filter sets and our highest-performance BrightLine filter sets.